UL Materials Testing and Certification (English)
Electric ignition source GWEPT - Glow Wire End Product Testing
This method is used to determine the flammability of end products using a glow wire loop. The test is suitable for end products which are exposed to an electrically heated source.
Standards for Electric ignition source GWEPT
UL 746A, IEC 60695-2-11, or equivalent standards
Electric ignition source GWFI - Glow Wire Flammability Testing
This method is used to determine the glow wire flammability index of a test specimen by means of a glow wire loop.
Standards for Electric ignition source
GWFIUL 746A, IEC 60695-2-12, or equivalent standards
Electric ignition source GWIT - Glow Wire Ignition Temperature
This method is used to determine the glow wire ignition temperature of a test specimen by means of a glow wire loop.
Standards for Electric ignition source GWIT
UL 746A, IEC 60695-2-13, or equivalent standards
Electric ignition source HWI - Hot Wire Ignition
This method is used to determine the ignition times of electrical insulation materials. The ignition time is the time which is required to ignite a wire-wound test specimen. This time serves as the assessment criterion for assigning the material to a Performance Level Class (PLC).
Standards for Electric ignition source HWI
UL 746A, ASTM D3874, or equivalent standards
- Test results for Electric ignition source GWEPT - Glow Wire End Product Testing
Pre-treatment:
24h / 15-35°C / 45-75 % relative humidityTest temperature in °C 500 550 600 650 700 750 800 850 900 960 Exposure time 30 s The specimen is considered to have passed the test if one of the following conditions is satisfied:
- The test specimen does not ignite
- The test specimen ignites but all of the following apply:
- The flames or incandescent burning (glow) are extinguished within 30 s of removing the glow wire, i.e. tE ≤ tA + 30 s
- The specified underlay for the test specimen does not ignite
- Test results for Electric ignition source GWFI - Glow Wire Flammability Testing
Pre-treatment:
48h / 23°C / 50 % relative humidityTest temperature in °C 550 600 650 700 750 800 850 900 960 Exposure time 30 s The specimen is considered to have passed the test if the following conditions are satisfied:
- The longest sustained flame or glow is extinguished within 30 s of removing the glow wire tR
- The specimen is not completely burned
- The tissue paper does not ignite
- Test results for Electric ignition source GWIT - Glow Wire Ignition Temperature
Pre-treatment:
48h / 23°C / 50 % rel. relative humidityTest temperature in °C 500 550 600 650 700 750 800 850 900 960 Exposure time 30 s The GWIT is 25 K (or 30 K) higher than the maximum test temperature according to this standardized method at which a tested material
- Does not ignite
- Ignites but where, in the event of sustained combustion with a flame, each incidence of ignition is sustained for a maximum of 5 s and the test specimen is not completely burned
- Test results for Electric ignition source HWI - Hot Wire Ignition
Pre-treatment:
> 48h / 23°C / 50 % relative humidityIgnition range (s) PLC > 120 0 > 60 to ≤ 120 1 > 30 to ≤ 60 2 > 15 to ≤ 30 3 > 7 to ≤ 15 4 0 to ≤ 7 5
Get connected with our sales team
Thanks for your interest in our products and services. Let's collect some information so we can connect you with the right person.